Test Description
The number(s) below refer to step(s) on the diagnostic table.
- (2) Applying 34 kPa (10 inches Hg) of vacuum to the MAP sensor should cause the voltage to change. Subtract the second voltage reading from the first. That voltage value should be more than 1.5 volts. When applying vacuum to the MAP sensor, the change in the voltage should happen instantly. A slow voltage change indicates a faulty MAP sensor.
- (3) Disconnect the MAP sensor from the bracket and twist the MAP sensor. Output changes more than 0.1 volt indicate a faulty connector or connection.
DIAGNOSTIC PROCEDURE
| Step | Action | Value(s) | Yes | No |
|---|---|---|---|---|
| 1 |
Is the difference in the two voltage readings less than the value specified? |
0.4 V | Go to Step 2 | Go to Step 5 |
| 2 |
Is the difference in voltage readings more than the value specified? |
1.5 V | System OK | Go to Step 3 |
| 3 | Inspect the MAP sensor connector terminals. Is the problem found? |
- | Go to Step 4 | Go to Step 5 |
| 4 | Repair the MAP sensor connector terminals as needed. Is the repair complete? |
- | System OK | - |
| 5 | Replace the MAP sensor. Is the repair complete? |
- | System OK | - |